Thermo-gravimetric analyser
The Perkin-Elmer TGA-7 thermogravimetric analyzer measures the weight loss of samples upon heating. The samples can be heated in a nitrogen or an air atmosphere, the gas being selectable via manual mode. The furnace can heat samples from room temperature to 900°C. The balance is sensitive to 0.1 microgram. The instrument uses Perkin-Elmer Pyris software for data acquisition and processing. |
|
|
|
||
Keywords: |
TGA, Perkin- Elmer 7 thermal analyzer, high temperature furnace. |
|
|
||
Measurements / Setups: |
Thermal loss, decomposition studies, weight % loss, activation energy. |
|
|
||
Specifications: |
Temperature range : 20-1000°C, Scanning rate ; 0.1 – 200 °C/min., Heating ,cooling and isothermal steps in one run, Temp precision : ± 2°C, Reproducibility to ± 2μg, Sensitivity: 0.1- 0.2 μg, Capacity: 1300 mg. |
|
|
||
Equipment: |
Platinum sample pan, sample thermocouple, hanging wire |
|
|
||
Publications: |
|
|
|
||
Location: |
ME 113 |
|
|
||
Research group: |
Elastomer Technology & Engineering (ETE) |
|
|
||
Contact: |
S. Maghami / W.K. Dierkes |
|